Current research projects

Image Measurement of insulated packaging
Image State of system and failure analyses
Image Solar Cooling
Image Ionocaloric cooling
Image Practical training, diploma, master, bachelor
Image CO₂ GAS HYDRATES FOR SUSTAINABLE ENERGY AND COOLING SOLUTIONS
Image Low Temperature Tribology
Image Reduction of primary noise sources of fans
Image Performance tests of condensing units
Image Air-flow test rig for fan characteristic measurement
Image Innovative Manufacturing Technologies for Cryosorption Systems
Image Certification of efficient air conditioning and ventilation systems through the new "indoor air quality seal" for non-residential buildings
Image In-Situ-Swelling Behaviour of Polymer Materials in Flammable Fluids
Image Preformance measurements of heat exchangers
Image Test rigs for refrigeration and heat pump technology
Image Industry 4.0 membrane heat and mass exchanger (i-MWÜ4.0)

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary