Current research projects

Image Helium extraction from natural gas
Image In-situ investigation concerning the swelling behaviour of polymer materials under elevated pressures and temperatures
Image IO-Scan - Integral measuring optical scanning method
Image Humidifier System for High-Purity Gases
Image Verification of storage suitability of cryo tubes
Image Measurements on ceiling mounted cooling systems
Image Industry 4.0 membrane heat and mass exchanger (i-MWÜ4.0)
Image Reduction of primary noise sources of fans
Image Calibration of Low Temperature Sensors
Image Testing of mobile leak detectors according to DIN EN 14624
Image Cryogenic liquid piston pumps for cold liquefied gases like LIN, LOX, LHe, LH2, LNG, LAr
Image Service offer for Leak Detection and Tightness Test
Image Testzentrum PLWP at ILK Dresden
Image Tribological investigations of oil-refrigerant-material-systems
Image Filter Tests
Image Tensile and compression testing

You are here:  Home /  Measurements and Tests


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary

Your Request

Further Projects - Measurements and Tests

Image

Mass Spectrometer

Determining the composition of gas mixtures in the high or ultra-high vacuum range

Image

Tensile and compression testing

Determination of yield strength, tensile strength and elongation at break

Image

Investigation of material-dependent parameters

Investigation of the permeation behavior

Image

Breakthrough Sensor for Adsorption Filters (BelA)

Sensor system for detecting an imminent breakthrough in gas filtration