Aktuelle Forschungsprojekte

Image Innovative small helium liquefier
Image Software for test rigs
Image Electrochemical decontamination of electrically conducting surfaces „EDeKo II“
Image Software for technical building equipment
Image High temperature heat pump
Image Optimizing HVAC operation with machine learning
Image Innovative Parahydrogen Generator Based on Magnets
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image Micro fluidic expansion valve
Image Ionocaloric cooling
Image Calibration of Low Temperature Sensors
Image Behavior of multiphase cryogenic fluids
Image Electrical components in refrigeration circuits
Image Intelligent innovative power supply for superconducting coils
Image Measurements on ceiling mounted cooling systems
Image Performance tests of refrigerant compressors

You are here:  Home /  Research and Development


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary