Aktuelle Forschungsprojekte

Image Test procedures for electrical components
Image Air-water heat pumps
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image Low Temperature Measuring Service
Image Certification of efficient air conditioning and ventilation systems through the new "indoor air quality seal" for non-residential buildings
Image Swirl-free on the move...
Image Service offer for Leak Detection and Tightness Test
Image Pulse-Tube-Refrigerator with sealed compressor
Image Multifunctional electronic modules for cryogenic applications
Image Investigation of coolants
Image Software for technical building equipment
Image Verification of storage suitability of cryo tubes
Image Calibration leak for the water bath leak test
Image Modular storage system for solar cooling
Image In-situ investigation concerning the swelling behaviour of polymer materials under elevated pressures and temperatures
Image Certifiable connection types in cryogenics

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary