Aktuelle Forschungsprojekte

Image Testing of mobile leak detectors according to DIN EN 14624
Image Cryogenic liquid piston pumps for cold liquefied gases like LIN, LOX, LHe, LH2, LNG, LAr
Image Energy efficiency consulting - cogeneration systems
Image Corrosion inhibitor for ammonia absorption systems
Image Low Temperature Tribology
Image Software for technical building equipment
Image Service offer for Leak Detection and Tightness Test
Image Low temperature – test facilities
Image 3D - Air flow sensor
Image Intelligent innovative power supply for superconducting coils
Image Optimizing HVAC operation with machine learning
Image Swirl-free on the move...
Image Electrical components in refrigeration circuits
Image Characterisation of Superconductors in Hydrogen Atmosphere
Image All-in-one device for freeze-drying and production of biomaterial
Image Cold meter

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary