Current research projects

Image Tensile and compression testing
Image Calibration leak for the water bath leak test
Image Humidifier System for High-Purity Gases
Image Refrigerants, lubricants and mixtures
Image State of system and failure analyses
Image Multifunctional electronic modules for cryogenic applications
Image Swirl-free on the move...
Image Energy efficiency consulting - cogeneration systems
Image Testzentrum PLWP at ILK Dresden
Image Software for test rigs
Image Certification of efficient air conditioning and ventilation systems through the new "indoor air quality seal" for non-residential buildings
Image Low Temperature Measuring Service
Image Overall System Optimization of Refrigeration Plant Systems for Energy Transition and Climate Protection
Image Measurements on ceiling mounted cooling systems
Image Mass Spectrometer
Image Optimizing HVAC operation with machine learning

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary