Aktuelle Forschungsprojekte

Image Range of services laboratory analyses
Image Corrosion inhibitor for ammonia absorption systems
Image Electrical components in refrigeration circuits
Image OVERALL SYSTEM OPTIMIZATION OF REFRIGERATION PLANT SYSTEMS FOR ENERGY TRANSITION AND CLIMATE PROTECTION
Image Behavior of multiphase cryogenic fluids
Image 3D - Air flow sensor
Image Influenced melting point of water by magnetic field
Image Laseroptical measurement
Image Development of test methods and test rigs for stationary integrated refrigeration units
Image Thermal engines
Image Software for technical building equipment
Image Cryogenic liquid piston pumps for cold liquefied gases like LIN, LOX, LHe, LH2, LNG, LAr
Image Investigation of materials
Image Non- invasive flow measurements
Image Electrochemical decontamination of electrically conducting surfaces „EDeKo II“
Image Multifunctional electronic modules for cryogenic applications

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary