Aktuelle Forschungsprojekte

Image Tribological investigations of oil-refrigerant-material-systems
Image In-situ investigation concerning the swelling behaviour of polymer materials under elevated pressures and temperatures
Image Low Temperature Measuring Service
Image Swirl-free on the move...
Image Corrosion inhibitor for ammonia absorption systems
Image Characterisation of Superconductors in Hydrogen Atmosphere
Image Test procedures for electrical components
Image Thermal engines
Image Service offer for Leak Detection and Tightness Test
Image Measurement of insulated packaging
Image Testzentrum PLWP at ILK Dresden
Image Ionocaloric cooling
Image High temperature heat pump
Image Behavior of multiphase cryogenic fluids
Image Combined building and system simulation
Image Cryostats, Non-Metallic and Metallic

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary

Your Request