Current research projects

Image Heat2Power
Image Verification of storage suitability of cryo tubes
Image Pulse-Tube-Refrigerator with sealed compressor
Image Testing of mobile leak detectors according to DIN EN 14624
Image Certification of efficient air conditioning and ventilation systems through the new "indoor air quality seal" for non-residential buildings
Image Industry 4.0 membrane heat and mass exchanger (i-MWÜ4.0)
Image Brine (water)-water heat pump
Image Preformance measurements of heat exchangers
Image Swirl-free on the move...
Image Range of services laboratory analyses
Image High Capacity Pulse Tube Cooler
Image Software modules
Image Tribological investigations of oil-refrigerant-material-systems
Image Cryostats, Non-Metallic and Metallic
Image Thermostatic Expansion Valves
Image Cryogenic liquid piston pumps for cold liquefied gases like LIN, LOX, LHe, LH2, LNG, LAr

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary