Current research projects

Image Corrosion inhibitor for ammonia absorption systems
Image State of system and failure analyses
Image Thermal engines
Image Computational fluid dynamics CFD
Image Intelligent innovative power supply for superconducting coils
Image Development of test methods and test rigs for stationary integrated refrigeration units
Image Practical training, diploma, master, bachelor
Image Investigation according to DIN EN ISO 14903
Image Ionocaloric cooling
Image Test method for high - temperature heat pump - oils
Image Measurement of insulated packaging
Image Electrochemical decontamination of electrically conducting surfaces „EDeKo II“
Image Performance tests of condensing units
Image Filter Tests
Image Software for technical building equipment
Image Innovative small helium liquefier

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary