Aktuelle Forschungsprojekte

Image Cryogenic liquid piston pumps for cold liquefied gases like LIN, LOX, LHe, LH2, LNG, LAr
Image Software for technical building equipment
Image Low temperature – test facilities
Image Test procedures for electrical components
Image Optimizing HVAC operation with machine learning
Image Solar Cooling
Image Micro heat exchangers in refrigeration
Image Energy efficiency consulting - cogeneration systems
Image All-in-one device for freeze-drying and production of biomaterial
Image Reducing the filling quantity
Image Thermostatic Expansion Valves
Image Helium extraction from natural gas
Image Industry 4.0 membrane heat and mass exchanger (i-MWÜ4.0)
Image High Capacity Pulse Tube Cooler
Image Ionocaloric cooling
Image Panel with indirect evaporative cooling via membrane

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary

Your Request

Further Projects

Image

Non- invasive flow measurements

PDPA - flow fields and particle sizes

Image

Computational fluid dynamics CFD

Scientific analysis of flows

Image

Humidifier System for High-Purity Gases

Nafion - moisture transfer

Image

Behavior of multiphase cryogenic fluids

experimental und numerical investigations