Current research projects

Image Industry 4.0 membrane heat and mass exchanger (i-MWÜ4.0)
Image ZeroHeatPump
Image Humidifier System for High-Purity Gases
Image Investigation of coolants
Image Performance tests of condensing units
Image Development of test methods and test rigs for stationary integrated refrigeration units
Image Certifiable connection types in cryogenics
Image Helium extraction from natural gas
Image Investigation according to DIN EN ISO 14903
Image Mass Spectrometer
Image Tensile and compression testing
Image Preformance measurements of heat exchangers
Image High temperature heat pump
Image Optimizing HVAC operation with machine learning
Image All-in-one device for freeze-drying and production of biomaterial
Image Tribological investigations of oil-refrigerant-material-systems

You are here:   /  Home


Cold meter

Industry

R. Paatzsch

+49-351-4081-5220

The fast way to refrigerating capacity

The ILK Dresden offers you special cold meters to determine the exact cooling capacity of your refrigeration system. These consist of a mass flow sensor, pressure sensors, temperature sensors and a suitable measured value acquisition. The associated software stores and analyses the measured values. The evaluation provides clear diagrams and key figures. Of course, the systems are calibrated before commissioning and can also be recalibrated.

Specification

  • refrigeration capacity range: ~ 0,5 kW … 300 kW (other ranges on request)
  • all common refrigerants
  • mass flow measurement is operating with the principle of a Coriolis flow meter
  • measurement of temperature and pressure in the refrigeration cycle
  • calculation of the refrigeration capacity from material properties using Reflib of the ILKs
  • comfortable evaluation and calculation software
  • monitoring of different parameters
  • hardware monitoring of measuring channels and network connection
  • error analysis

Your Request

Further Projects