Current research projects

Image Multifunctional electronic modules for cryogenic applications
Image Service offer for Leak Detection and Tightness Test
Image Intelligent innovative power supply for superconducting coils
Image Investigation according to DIN EN ISO 14903
Image Mass Spectrometer
Image Calibration leak for the water bath leak test
Image Industry 4.0 membrane heat and mass exchanger (i-MWÜ4.0)
Image Hydrogen and methane testing field at the ILK
Image Heat2Power
Image Certification of efficient air conditioning and ventilation systems through the new "indoor air quality seal" for non-residential buildings
Image Test rigs for refrigeration and heat pump technology
Image Hybrid- Fluid for CO2-Sublimation Cycle
Image Non- invasive flow measurements
Image Testzentrum PLWP at ILK Dresden
Image Electrochemical decontamination of electrically conducting surfaces „EDeKo II“
Image Cool Up

You are here:  Home /  Measurements and Tests


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary