Current research projects

Image Micro heat exchangers in refrigeration
Image Low temperature – test facilities
Image Influenced melting point of water by magnetic field
Image Computational fluid dynamics CFD
Image Testing of mobile leak detectors according to DIN EN 14624
Image Air-flow test rig for fan characteristic measurement
Image Service offer for Leak Detection and Tightness Test
Image State of system and failure analyses
Image Software for technical building equipment
Image Reduction of primary noise sources of fans
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image Low Temperature Tribology
Image Investigation of material-dependent parameters
Image Cryostats, Non-Metallic and Metallic
Image Modular storage system for solar cooling
Image Industry 4.0 membrane heat and mass exchanger (i-MWÜ4.0)

You are here:   /  Home


Humidifier System for High-Purity Gases

Industry and Research Instituts

Moritz Kuhn

+49-351-4081-5120

Nafion - moisture transfer

At the ILK Dresden a humidification system was developed and manufactured which is especially suitable for research facilities where very low dew point temperatures have to be adjusted and at the same time a high gas purity has to be guaranteed. The core component is a PermaPure® humidifier with Nafion® moisture exchange. The humidification system precisely controls the dew point via a bypass mixing with dry gas and also takes over the automatic filling of evacuated containers. The control system is optimised to prevent recondensation. Humidification can be controlled from any plant system via an interface. Table 1 shows the technical data of a humidification system for the humidification of synthetic air realized for the KIT in Karlsruhe. The system can be adapted for other applications, and all parameters can be adjusted to the respective requirements.

Table  1: Technical Data Humidifier System
Parameter Value
setpoint gas flow 750 litre/min
maximum relative humidity 95%
minimum dew point -60°C / -76°F
degree of gas purity 5.0
gas purity, number of particle < 5 particle/cm³