Aktuelle Forschungsprojekte

Image Development of test methods and test rigs for stationary integrated refrigeration units
Image Thermal engines
Image Reduction of primary noise sources of fans
Image Test procedures for electrical components
Image Pulse-Tube-Refrigerator with sealed compressor
Image Optimizing HVAC operation with machine learning
Image State of system and failure analyses
Image Ice Slurry Generation
Image Innovative Parahydrogen Generator Based on Magnets
Image Refrigerants, lubricants and mixtures
Image IO-Scan - Integral measuring optical scanning method
Image Micro fluidic expansion valve
Image Cool Up
Image In-situ investigation concerning the swelling behaviour of polymer materials under elevated pressures and temperatures
Image Verification of storage suitability of cryo tubes
Image Cryostats, Non-Metallic and Metallic

You are here:   /  Home


Humidifier System for High-Purity Gases

Industry and Research Instituts

Moritz Kuhn

+49-351-4081-5120

Nafion - moisture transfer

At the ILK Dresden a humidification system was developed and manufactured which is especially suitable for research facilities where very low dew point temperatures have to be adjusted and at the same time a high gas purity has to be guaranteed. The core component is a PermaPure® humidifier with Nafion® moisture exchange. The humidification system precisely controls the dew point via a bypass mixing with dry gas and also takes over the automatic filling of evacuated containers. The control system is optimised to prevent recondensation. Humidification can be controlled from any plant system via an interface. Table 1 shows the technical data of a humidification system for the humidification of synthetic air realized for the KIT in Karlsruhe. The system can be adapted for other applications, and all parameters can be adjusted to the respective requirements.

Table  1: Technical Data Humidifier System
Parameter Value
setpoint gas flow 750 litre/min
maximum relative humidity 95%
minimum dew point -60°C / -76°F
degree of gas purity 5.0
gas purity, number of particle < 5 particle/cm³