Current research projects

Image Cool Up
Image Humidifier System for High-Purity Gases
Image Software for technical building equipment
Image Development of test methods and test rigs for stationary integrated refrigeration units
Image Preformance measurements of heat exchangers
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image Electrical components in refrigeration circuits
Image Testzentrum PLWP at ILK Dresden
Image Certifiable connection types in cryogenics
Image Swirl-free on the move...
Image Reducing the filling quantity
Image Measurement of insulated packaging
Image In-Situ-Swelling Behaviour of Polymer Materials in Flammable Fluids
Image Testing of mobile leak detectors according to DIN EN 14624
Image Laseroptical measurement
Image Verification of storage suitability of cryo tubes

You are here:   /  Home


Low noise and non metallic liquid-helium cryostat

Industry and Research Institutes

Dr. rer. nat. Andreas Kade

+49-351-4081-5117

Low-noise Magnetic Field Cryostat for SQUID-Applications

Technical Data for the Cryostats

parametervalue
volume of liquid helium5 to 12 litre (other volume possible)
time to complete evaporation of helium3 to 7 days (depends on size)
heat load0.1 W (in standby operation)
rate of helium evaporation≤ 3 litre / day
holding time of the liquid heliumup to 4 days without refilling
initial noise of the cryostat< 3 fT / Hz½
helium leak test (He-vessel)< 1 × 10-11 mbar l / s

ILK Dresden developed low noise magnetic field and non metallic helium cryostats made of GRP (glass reinforced plastics) with a high helium and vacuum holding time.

The cryostats can operate in a direction independent operation mode (tiltable arrangement) and hence they are suitable for a plurality of several applications.

The ILK concept offers much lower permeation rates inside the He-reservoir as comparable models from other manufacturers and is therefore perfect for SQUID (superconducting quantum interference device) applications and other long-term measurements.
 
The technical design of the cryostats is thus predestined for future sensor generation.

Specification

  • suitable for the cooling of SQUID-Sensors
  • non-metallic (GRP)
  • fast filling with LHe via thermosiphon
  • GRP exhaust gas cooled radiation shield for high efficiency.
  • small cold-warm distance feasible
  • low maintenance 
  • other design variants possible on costumer request

Low noise magnetic field

The low-noise magnetic field of the cryostat was tested in the magnetically shielded room BMSR-1 of PTB Berlin and is smaller than the lowest resolution limit of the measurement system used, see figure below.


Your Request

Further Projects

Image

ZeroHeatPump

Performance management of small heat pumps without energy consumption

Image

Mass Spectrometer

Determining the composition of gas mixtures in the high or ultra-high vacuum range