Current research projects

Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image Investigation of material-dependent parameters
Image ZeroHeatPump
Image Air-water heat pumps
Image Range of services laboratory analyses
Image IO-Scan - Integral measuring optical scanning method
Image Cryostats, Non-Metallic and Metallic
Image High Capacity Pulse Tube Cooler
Image Breakthrough Sensor for Adsorption Filters (BelA)
Image Non- invasive flow measurements
Image Cryogenic liquid piston pumps for cold liquefied gases like LIN, LOX, LHe, LH2, LNG, LAr
Image Investigation of materials
Image Software for technical building equipment
Image Energy efficiency consulting - cogeneration systems
Image Micro heat exchangers in refrigeration
Image Optimizing HVAC operation with machine learning

You are here:   /  Home


Tensile and compression testing

Industry and research institutes

Gunar Schroeder

+49-351-4081-5129

Determination of yield strength, tensile strength and elongation at break

A tensile or compression test is a quasi-static, destructive test method for determining material characteristics such as yield strength, tensile strength and elongation at break. The specimen form and the procedure of such tests are standardized. The ISO 9862 standard defines the exact procedure and the measured variables to be recorded for metallic materials and the DIN 50125 standard defines the specimen shape and its production method. In addition, the arrangement can generally be used to record force-displacement curves with high accuracy.

To carry out the test, the specimens are clamped centrally in the grippers or positioned centrally between the compression plates. In the tensile test, the specimens are pulled apart in a defined manner. In the compression test, on the other hand, a compressive force is applied to the test specimen, e.g. rectangular prisms, cylinders or tubes, in a uniaxial stress state. In both tests, the force and change in length are recorded continuously and evaluated graphically. On request, the tensile and compression test can be carried out at very low temperatures.

Parameter Limit values samples
Materials foils, plastics, composites, metals
Typical dimension of plastic samples Lר: 200×10 mm (7.8"×0.4") or
L×W×H: 200×10×10 mm (7.8"×0.4"×0.4")
Measurement range of force and displacement ±5 kN (±1124 lbf), 400 mm (15.8")
Standardized testing force-distance, force-time, tension-extension

 


Your Request