Current research projects

Image Verification of storage suitability of cryo tubes
Image Cool Up
Image Ionocaloric cooling
Image Measurements on ceiling mounted cooling systems
Image Humidifier System for High-Purity Gases
Image Swirl-free on the move...
Image Test method for high - temperature heat pump - oils
Image Heat2Power
Image Software modules
Image Electrical components in refrigeration circuits
Image Test rigs for refrigeration and heat pump technology
Image Characterisation of Superconductors in Hydrogen Atmosphere
Image Corrosion inhibitor for ammonia absorption systems
Image Low Temperature Measuring Service
Image Certification of efficient air conditioning and ventilation systems through the new "indoor air quality seal" for non-residential buildings
Image Computational fluid dynamics CFD

You are here:  Home /  Measurements and Tests


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary

Your Request

Further Projects - Measurements and Tests