Current research projects

Image Micro heat exchangers in refrigeration
Image Test rigs for refrigeration and heat pump technology
Image Electrical components in refrigeration circuits
Image Industry 4.0 membrane heat and mass exchanger (i-MWÜ4.0)
Image Filter Tests
Image Optimizing HVAC operation with machine learning
Image Brine (water)-water heat pump
Image Intelligent innovative power supply for superconducting coils
Image Testing of mobile leak detectors according to DIN EN 14624
Image Ionocaloric cooling
Image Mass Spectrometer
Image Performance tests of refrigerant compressors
Image Multifunctional electronic modules for cryogenic applications
Image Ice Slurry Generation
Image Production of novel barrier layers on polymer materials to reduce hydrogen permeation
Image Performance tests of condensing units

You are here:  Home /  Measurements and Tests


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary

Your Request