Current research projects

Image Production of novel barrier layers on polymer materials to reduce hydrogen permeation
Image Cryostats, Non-Metallic and Metallic
Image Performance tests of condensing units
Image Panel with indirect evaporative cooling via membrane
Image IO-Scan - Integral measuring optical scanning method
Image ZeroHeatPump
Image Test method for high - temperature heat pump - oils
Image Lifetime prediction of hermetic compressor systems
Image Overall System Optimization of Refrigeration Plant Systems for Energy Transition and Climate Protection
Image Intelligent innovative power supply for superconducting coils
Image High temperature heat pump
Image Breakthrough Sensor for Adsorption Filters (BelA)
Image Cold meter
Image Non- invasive flow measurements
Image Ice Slurry Generation
Image Testzentrum PLWP at ILK Dresden

You are here:  Home /  Measurements and Tests


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary

Your Request

Further Projects - Measurements and Tests