Current research projects

Image Practical training, diploma, master, bachelor
Image Reducing the filling quantity
Image Investigation of materials
Image Heat2Power
Image Thermostatic Expansion Valves
Image Measurements on ceiling mounted cooling systems
Image Cold meter
Image Lifetime prediction of hermetic compressor systems
Image Certifiable connection types in cryogenics
Image Test procedures for electrical components
Image Combined building and system simulation
Image Optimizing HVAC operation with machine learning
Image Low noise and non metallic liquid-helium cryostat
Image Low Temperature Measuring Service
Image IO-Scan - Integral measuring optical scanning method
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K

You are here:  Home /  Measurements and Tests


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary