Current research projects

Image Laseroptical measurement
Image Innovative small helium liquefier
Image Certification of efficient air conditioning and ventilation systems through the new "indoor air quality seal" for non-residential buildings
Image Heat2Power
Image Measurement of insulated packaging
Image Electrochemical decontamination of electrically conducting surfaces „EDeKo II“
Image Non- invasive flow measurements
Image High Capacity Pulse Tube Cooler
Image Micro fluidic expansion valve
Image Behavior of multiphase cryogenic fluids
Image Helium extraction from natural gas
Image Investigation of material-dependent parameters
Image Investigation according to DIN EN ISO 14903
Image Calibration leak for the water bath leak test
Image Electrical components in refrigeration circuits
Image In-Situ-Swelling Behaviour of Polymer Materials in Flammable Fluids

You are here:  Home /  Measurements and Tests


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary