Current research projects

Image Cold meter
Image Innovative Parahydrogen Generator Based on Magnets
Image Certification of efficient air conditioning and ventilation systems through the new "indoor air quality seal" for non-residential buildings
Image Brine (water)-water heat pump
Image Cool Up
Image Reduction of primary noise sources of fans
Image Electrochemical decontamination of electrically conducting surfaces „EDeKo II“
Image Measurements on ceiling mounted cooling systems
Image Helium extraction from natural gas
Image Electrical components in refrigeration circuits
Image Development of test methods and test rigs for stationary integrated refrigeration units
Image Range of services laboratory analyses
Image Humidifier System for High-Purity Gases
Image Micro heat exchangers in refrigeration
Image Ice Slurry Generation
Image Characterisation of Superconductors in Hydrogen Atmosphere

You are here:  Home /  Measurements and Tests


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary

Your Request

Further Projects - Measurements and Tests

Image

Preformance measurements of heat exchangers

Is the heat exchanger properly sized?

Image

Filter Tests

INDUSTRIAL AND LABORATORY PRECIPITATORS

Image

Investigation of coolants

Secondary loop refrigerants

Image

Refrigerants, lubricants and mixtures

Determination of working fluid properties