Current research projects

Image CFE-Test of Cooker Hoods
Image Measurement of insulated packaging
Image Pulse-Tube-Refrigerator with sealed compressor
Image Corrosion inhibitor for ammonia absorption systems
Image Innovative Manufacturing Technologies for Cryosorption Systems
Image Air-flow test rig for fan characteristic measurement
Image Electrochemical decontamination of electrically conducting surfaces „EDeKo II“
Image Filter Tests
Image 3D - Air flow sensor
Image In-situ investigation concerning the swelling behaviour of polymer materials under elevated pressures and temperatures
Image Non- invasive flow measurements
Image Development of a Cryogenic Magnetic Air Separation Unit
Image Helium extraction from natural gas
Image Tribological investigations of oil-refrigerant-material-systems
Image Refrigerants, lubricants and mixtures
Image Service offer for Leak Detection and Tightness Test

You are here:  Home /  Measurements and Tests


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary

Your Request

Further Projects - Measurements and Tests

Image

Mass Spectrometer

Determining the composition of gas mixtures in the high or ultra-high vacuum range

Image

Tensile and compression testing

Determination of yield strength, tensile strength and elongation at break

Image

Investigation of material-dependent parameters

Investigation of the permeation behavior

Image

Breakthrough Sensor for Adsorption Filters (BelA)

Sensor system for detecting an imminent breakthrough in gas filtration