Current research projects

Image Investigation of coolants
Image Innovative small helium liquefier
Image Development of test methods and test rigs for stationary integrated refrigeration units
Image Performance tests of condensing units
Image Cryogenic liquid piston pumps for cold liquefied gases like LIN, LOX, LHe, LH2, LNG, LAr
Image Practical training, diploma, master, bachelor
Image Brine (water)-water heat pump
Image Computational fluid dynamics CFD
Image Non- invasive flow measurements
Image Low temperature – test facilities
Image Cryostats, Non-Metallic and Metallic
Image Calibration leak for the water bath leak test
Image Electrochemical decontamination of electrically conducting surfaces „EDeKo II“
Image Software for technical building equipment
Image IO-Scan - Integral measuring optical scanning method
Image Refrigerants, lubricants and mixtures

You are here:  Home /  Measurements and Tests


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary