Current research projects

Image Heat2Power
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image CFE-Test of Cooker Hoods
Image Hybrid- Fluid for CO2-Sublimation Cycle
Image Industry 4.0 membrane heat and mass exchanger (i-MWÜ4.0)
Image Micro heat exchangers in refrigeration
Image Software modules
Image Measurements on ceiling mounted cooling systems
Image Laseroptical measurement
Image Energy efficiency consulting - cogeneration systems
Image Humidifier System for High-Purity Gases
Image Investigation of material-dependent parameters
Image Air-water heat pumps
Image Reduction of primary noise sources of fans
Image High Capacity Pulse Tube Cooler
Image All-in-one device for freeze-drying and production of biomaterial

You are here:  Home /  Measurements and Tests


Mass Spectrometer

Industrie und Forschungsinstitute

Gunar Schroeder

+49-351-4081-5129

Determining the composition of gas mixtures in the high or ultra-high vacuum range

Mass spectrometry is a method for determining the composition of gas mixtures in the high or ultra-high vacuum range. It can be used to carry out qualitative and quantitative gas analyses and leak detection or to determine trace contamination.

The individual neutral gas particles are ionized by an electron impact ion source and separated from each other in an electric quadrupole field according to their mass-to-charge ratio. The filtered ions are then recorded by a detector. The ion current detected is proportional to the partial pressure of the respective gas component. For a more precise quantitative gas analysis, the relatively operating measuring instrument can be calibrated with a calibration gas.

The test set-up consists of a test chamber with a connected turbo-molecular-pump and fore vacuum pump. The mass spectrometer requires a vacuum <1,5×10-7 psi (<1×10-5 mbar) to be able to measure with high precision. The mass spectrometer and a vacuum sensor are mounted on the test chamber. The mass spectrometer is connected to a measuring PC with corresponding software.

Available measurement modes are:

  • Vacuum analysis, to determine residual gas components, evaluation as a plot of ion current and mass
  • Leak detection, with helium or another tracer gas, the ion current of the tracer gas or the corresponding mass is recorded and plotted over the time
  • Mass spectrum, evaluation of ion current and mass with user-defined specifications and time trend
Paramter Test value
Analysis Vacuum analysis, leak detection, mass spectrum, trace contamination
Detectable gases H2, He, N2, H2O, O2, CO2 usw.
Mass range 1 – 100 Da (1,66×10-27 – 1,66×10-25 kg)
Detection limit 1.5×10-15 psi / 3×10-13 mbar
Min. vacuum pressure and max. temperature range <7.3×10-6 psi / 5×10-4 mbar, < 150°C / 302°F

[Translate to EN:]


Your Request

Further Projects - Measurements and Tests

Image

Preformance measurements of heat exchangers

Is the heat exchanger properly sized?

Image

Filter Tests

INDUSTRIAL AND LABORATORY PRECIPITATORS

Image

Investigation of coolants

Secondary loop refrigerants

Image

Refrigerants, lubricants and mixtures

Determination of working fluid properties