Current research projects

Image Practical training, diploma, master, bachelor
Image Development of a Cryogenic Magnetic Air Separation Unit
Image Tribological investigations of oil-refrigerant-material-systems
Image Measurement of insulated packaging
Image Non- invasive flow measurements
Image Refrigerants, lubricants and mixtures
Image Helium extraction from natural gas
Image Production of novel barrier layers on polymer materials to reduce hydrogen permeation
Image ZeroHeatPump
Image Reduction of primary noise sources of fans
Image Thermostatic Expansion Valves
Image Test rigs for refrigeration and heat pump technology
Image Certifiable connection types in cryogenics
Image Micro heat exchangers in refrigeration
Image All-in-one device for freeze-drying and production of biomaterial
Image Swirl-free on the move...

You are here:  Home /  Measurements and Tests


Mass Spectrometer

Industrie und Forschungsinstitute

Gunar Schroeder

+49-351-4081-5129

Determining the composition of gas mixtures in the high or ultra-high vacuum range

Mass spectrometry is a method for determining the composition of gas mixtures in the high or ultra-high vacuum range. It can be used to carry out qualitative and quantitative gas analyses and leak detection or to determine trace contamination.

The individual neutral gas particles are ionized by an electron impact ion source and separated from each other in an electric quadrupole field according to their mass-to-charge ratio. The filtered ions are then recorded by a detector. The ion current detected is proportional to the partial pressure of the respective gas component. For a more precise quantitative gas analysis, the relatively operating measuring instrument can be calibrated with a calibration gas.

The test set-up consists of a test chamber with a connected turbo-molecular-pump and fore vacuum pump. The mass spectrometer requires a vacuum <1,5×10-7 psi (<1×10-5 mbar) to be able to measure with high precision. The mass spectrometer and a vacuum sensor are mounted on the test chamber. The mass spectrometer is connected to a measuring PC with corresponding software.

Available measurement modes are:

  • Vacuum analysis, to determine residual gas components, evaluation as a plot of ion current and mass
  • Leak detection, with helium or another tracer gas, the ion current of the tracer gas or the corresponding mass is recorded and plotted over the time
  • Mass spectrum, evaluation of ion current and mass with user-defined specifications and time trend
Paramter Test value
Analysis Vacuum analysis, leak detection, mass spectrum, trace contamination
Detectable gases H2, He, N2, H2O, O2, CO2 usw.
Mass range 1 – 100 Da (1,66×10-27 – 1,66×10-25 kg)
Detection limit 1.5×10-15 psi / 3×10-13 mbar
Min. vacuum pressure and max. temperature range <7.3×10-6 psi / 5×10-4 mbar, < 150°C / 302°F

[Translate to EN:]


Your Request

Further Projects - Measurements and Tests

Image

Non- invasive flow measurements

PDPA - flow fields and particle sizes

Image

Laseroptical measurement

PIV and LDA / PDA

Image

Performance tests of condensing units

Does your condensing unit perform well?

Image

State of system and failure analyses

Cause of the failure unknown?