Current research projects

Image Investigation of material-dependent parameters
Image Cool Up
Image Behavior of multiphase cryogenic fluids
Image Overall System Optimization of Refrigeration Plant Systems for Energy Transition and Climate Protection
Image Low Temperature Tribology
Image Development of test methods and test rigs for stationary integrated refrigeration units
Image Combined building and system simulation
Image Solar Cooling
Image Hydrogen and methane testing field at the ILK
Image Influenced melting point of water by magnetic field
Image 3D - Air flow sensor
Image Breakthrough Sensor for Adsorption Filters (BelA)
Image Practical training, diploma, master, bachelor
Image Test method for high - temperature heat pump - oils
Image Innovative Parahydrogen Generator Based on Magnets
Image Range of services laboratory analyses

You are here:   /  Home


Calibration of Low Temperature Sensors

141,142

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary