Current research projects

Image Humidifier System for High-Purity Gases
Image In-Situ-Swelling Behaviour of Polymer Materials in Flammable Fluids
Image Lifetime prediction of hermetic compressor systems
Image Certification of efficient air conditioning and ventilation systems through the new "indoor air quality seal" for non-residential buildings
Image Low noise and non metallic liquid-helium cryostat
Image 3D - Air flow sensor
Image Investigation of material-dependent parameters
Image Verification of storage suitability of cryo tubes
Image Industry 4.0 membrane heat and mass exchanger (i-MWÜ4.0)
Image Test procedures for electrical components
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image Hybrid- Fluid for CO2-Sublimation Cycle
Image High Capacity Pulse Tube Cooler
Image Cold meter
Image Calibration leak for the water bath leak test
Image Innovative Parahydrogen Generator Based on Magnets

You are here:   /  Home


Calibration of Low Temperature Sensors

141,142

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary

Your Request