Current research projects

Image Tensile and compression testing
Image Ionocaloric cooling
Image Characterisation of Superconductors in Hydrogen Atmosphere
Image Tribological investigations of oil-refrigerant-material-systems
Image In-situ investigation concerning the swelling behaviour of polymer materials under elevated pressures and temperatures
Image Testzentrum PLWP at ILK Dresden
Image Multifunctional electronic modules for cryogenic applications
Image Cold meter
Image Panel with indirect evaporative cooling via membrane
Image Intelligent innovative power supply for superconducting coils
Image Heat2Power
Image Innovative Parahydrogen Generator Based on Magnets
Image 3D - Air flow sensor
Image Modular storage system for solar cooling
Image Thermostatic Expansion Valves
Image Performance tests of refrigerant compressors

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary