Current research projects

Image Tensile and compression testing
Image Intelligent innovative power supply for superconducting coils
Image In-situ investigation concerning the swelling behaviour of polymer materials under elevated pressures and temperatures
Image Breakthrough Sensor for Adsorption Filters (BelA)
Image Multifunctional electronic modules for cryogenic applications
Image Swirl-free on the move...
Image Measurements on ceiling mounted cooling systems
Image Calibration of Low Temperature Sensors
Image Practical training, diploma, master, bachelor
Image Investigation of material-dependent parameters
Image Filter Tests
Image In-Situ-Swelling Behaviour of Polymer Materials in Flammable Fluids
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image Range of services laboratory analyses
Image Low temperature – test facilities
Image Testing of mobile leak detectors according to DIN EN 14624

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary