Current research projects

Image Calibration of Low Temperature Sensors
Image Investigation of materials
Image Performance tests of refrigerant compressors
Image Hydrogen and methane testing field at the ILK
Image Service offer for Leak Detection and Tightness Test
Image Thermal engines
Image Low temperature – test facilities
Image High Capacity Pulse Tube Cooler
Image Optimizing HVAC operation with machine learning
Image Lifetime prediction of hermetic compressor systems
Image Multifunctional electronic modules for cryogenic applications
Image Test rigs for refrigeration and heat pump technology
Image Production of novel barrier layers on polymer materials to reduce hydrogen permeation
Image Ionocaloric cooling
Image Electrochemical decontamination of electrically conducting surfaces „EDeKo II“
Image Brine (water)-water heat pump

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary