Current research projects

Image Energy efficiency consulting - cogeneration systems
Image Development of a Cryogenic Magnetic Air Separation Unit
Image Low Temperature Measuring Service
Image Practical training, diploma, master, bachelor
Image Cold meter
Image Certification of efficient air conditioning and ventilation systems through the new "indoor air quality seal" for non-residential buildings
Image Intelligent innovative power supply for superconducting coils
Image Humidifier System for High-Purity Gases
Image Measurement of insulated packaging
Image Verification of storage suitability of cryo tubes
Image Hydrogen and methane testing field at the ILK
Image All-in-one device for freeze-drying and production of biomaterial
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image Service offer for Leak Detection and Tightness Test
Image Behavior of multiphase cryogenic fluids
Image Non- invasive flow measurements

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary