Aktuelle Forschungsprojekte

Image Service offer for Leak Detection and Tightness Test
Image Filter Tests
Image Test procedures for electrical components
Image Ice Slurry Generation
Image Combined building and system simulation
Image Helium extraction from natural gas
Image Lifetime prediction of hermetic compressor systems
Image Practical training, diploma, master, bachelor
Image Innovative Parahydrogen Generator Based on Magnets
Image Measurements on ceiling mounted cooling systems
Image Corrosion inhibitor for ammonia absorption systems
Image Development of test methods and test rigs for stationary integrated refrigeration units
Image Electrochemical decontamination of electrically conducting surfaces „EDeKo II“
Image Influenced melting point of water by magnetic field
Image Software modules
Image Certification of efficient air conditioning and ventilation systems through the new "indoor air quality seal" for non-residential buildings

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary

Your Request

Further Projects

Image

Electrical components in refrigeration circuits

High voltage tests under real conditions

Image

Influenced melting point of water by magnetic field

Controlled sub-cooling of products in freezing processes

Image

Investigation of materials

Investigations regarding the compatibility of materials with refrigerants, oils and heat transfer fluids

Image