Current research projects

Image Cryostats, Non-Metallic and Metallic
Image Measurements on ceiling mounted cooling systems
Image Preformance measurements of heat exchangers
Image Combined building and system simulation
Image Characterisation of Superconductors in Hydrogen Atmosphere
Image State of system and failure analyses
Image Investigation according to DIN EN ISO 14903
Image Practical training, diploma, master, bachelor
Image Micro heat exchangers in refrigeration
Image Development of a Cryogenic Magnetic Air Separation Unit
Image Electrical components in refrigeration circuits
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image Heat2Power
Image Low noise and non metallic liquid-helium cryostat
Image Thermal engines
Image Micro fluidic expansion valve

You are here:   /  Home


Mass Spectrometer

Industrie und Forschungsinstitute

Gunar Schroeder

+49-351-4081-5129

Determining the composition of gas mixtures in the high or ultra-high vacuum range

Mass spectrometry is a method for determining the composition of gas mixtures in the high or ultra-high vacuum range. It can be used to carry out qualitative and quantitative gas analyses and leak detection or to determine trace contamination.

The individual neutral gas particles are ionized by an electron impact ion source and separated from each other in an electric quadrupole field according to their mass-to-charge ratio. The filtered ions are then recorded by a detector. The ion current detected is proportional to the partial pressure of the respective gas component. For a more precise quantitative gas analysis, the relatively operating measuring instrument can be calibrated with a calibration gas.

The test set-up consists of a test chamber with a connected turbo-molecular-pump and fore vacuum pump. The mass spectrometer requires a vacuum <1,5×10-7 psi (<1×10-5 mbar) to be able to measure with high precision. The mass spectrometer and a vacuum sensor are mounted on the test chamber. The mass spectrometer is connected to a measuring PC with corresponding software.

Available measurement modes are:

  • Vacuum analysis, to determine residual gas components, evaluation as a plot of ion current and mass
  • Leak detection, with helium or another tracer gas, the ion current of the tracer gas or the corresponding mass is recorded and plotted over the time
  • Mass spectrum, evaluation of ion current and mass with user-defined specifications and time trend
Paramter Test value
Analysis Vacuum analysis, leak detection, mass spectrum, trace contamination
Detectable gases H2, He, N2, H2O, O2, CO2 usw.
Mass range 1 – 100 Da (1,66×10-27 – 1,66×10-25 kg)
Detection limit 1.5×10-15 psi / 3×10-13 mbar
Min. vacuum pressure and max. temperature range <7.3×10-6 psi / 5×10-4 mbar, < 150°C / 302°F

[Translate to EN:]


Your Request

Further Projects

Image

Characterisation of Superconductors in Hydrogen Atmosphere

Are superconductors really compatible with hydrogen?

Image

Corrosion inhibitor for ammonia absorption systems

An alternative to chromium(VI) compounds

Image

Development of a Cryogenic Magnetic Air Separation Unit

Oxygen Enrichment by Applied Cryogenic Magnetohydrodynamics

Image

Software for test rigs

Individual software for complex tests and evaluation

Image

Brine (water)-water heat pump

Test according DIN EN 14511 and 14825