Current research projects

Image Mass Spectrometer
Image Panel with indirect evaporative cooling via membrane
Image Development of a Cryogenic Magnetic Air Separation Unit
Image Thermal engines
Image Non- invasive flow measurements
Image Cool Up
Image Hybrid- Fluid for CO2-Sublimation Cycle
Image In-Situ-Swelling Behaviour of Polymer Materials in Flammable Fluids
Image All-in-one device for freeze-drying and production of biomaterial
Image Testing of mobile leak detectors according to DIN EN 14624
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image Calibration leak for the water bath leak test
Image Multifunctional electronic modules for cryogenic applications
Image Low Temperature Tribology
Image Tensile and compression testing
Image ZeroHeatPump

You are here:  Home /  Measurements and Tests


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary