Current research projects

Image Reducing the filling quantity
Image Performance tests of condensing units
Image Solar Cooling
Image Service offer for Leak Detection and Tightness Test
Image Low temperature – test facilities
Image Humidifier System for High-Purity Gases
Image Investigation of coolants
Image Industry 4.0 membrane heat and mass exchanger (i-MWÜ4.0)
Image Low Temperature Tribology
Image Multifunctional electronic modules for cryogenic applications
Image Performance tests of refrigerant compressors
Image Swirl-free on the move...
Image Software modules
Image Certification of efficient air conditioning and ventilation systems through the new "indoor air quality seal" for non-residential buildings
Image Ice Slurry Generation
Image Reduction of primary noise sources of fans

You are here:  Home /  Measurements and Tests


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary