Current research projects

Image Mass Spectrometer
Image Cryostats, Non-Metallic and Metallic
Image Behavior of multiphase cryogenic fluids
Image Performance tests of condensing units
Image Modular storage system for solar cooling
Image Verification of storage suitability of cryo tubes
Image Multifunctional electronic modules for cryogenic applications
Image State of system and failure analyses
Image Certification of efficient air conditioning and ventilation systems through the new "indoor air quality seal" for non-residential buildings
Image Development of a Cryogenic Magnetic Air Separation Unit
Image Innovative Manufacturing Technologies for Cryosorption Systems
Image Low noise and non metallic liquid-helium cryostat
Image In-Situ-Swelling Behaviour of Polymer Materials in Flammable Fluids
Image Cold meter
Image Measurement of insulated packaging
Image Intelligent innovative power supply for superconducting coils

You are here:  Home /  Research and Development


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary