Current research projects

Image Test rigs for refrigeration and heat pump technology
Image Investigation according to DIN EN ISO 14903
Image Humidifier System for High-Purity Gases
Image Helium extraction from natural gas
Image ZeroHeatPump
Image Characterisation of Superconductors in Hydrogen Atmosphere
Image Pulse-Tube-Refrigerator with sealed compressor
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image Behavior of multiphase cryogenic fluids
Image Innovative Parahydrogen Generator Based on Magnets
Image Corrosion inhibitor for ammonia absorption systems
Image Performance tests of condensing units
Image Testzentrum PLWP at ILK Dresden
Image Innovative Manufacturing Technologies for Cryosorption Systems
Image Cryogenic liquid piston pumps for cold liquefied gases like LIN, LOX, LHe, LH2, LNG, LAr
Image CO₂ GAS HYDRATES FOR SUSTAINABLE ENERGY AND COOLING SOLUTIONS

You are here:  Home /  Research and Development


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary