Current research projects

Image Reduction of primary noise sources of fans
Image Modular storage system for solar cooling
Image Low noise and non metallic liquid-helium cryostat
Image Reducing the filling quantity
Image Development of test methods and test rigs for stationary integrated refrigeration units
Image Performance tests of refrigerant compressors
Image Investigation of coolants
Image ZeroHeatPump
Image Lifetime prediction of hermetic compressor systems
Image In-situ investigation concerning the swelling behaviour of polymer materials under elevated pressures and temperatures
Image Software modules
Image CO₂ GAS HYDRATES FOR SUSTAINABLE ENERGY AND COOLING SOLUTIONS
Image Micro heat exchangers in refrigeration
Image Electrochemical decontamination of electrically conducting surfaces „EDeKo II“
Image IO-Scan - Integral measuring optical scanning method
Image Calibration of Low Temperature Sensors

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary