Current research projects

Image Reduction of primary noise sources of fans
Image Certification of efficient air conditioning and ventilation systems through the new "indoor air quality seal" for non-residential buildings
Image Air-flow test rig for fan characteristic measurement
Image High Capacity Pulse Tube Cooler
Image Modular storage system for solar cooling
Image Investigation of material-dependent parameters
Image Reducing the filling quantity
Image Solar Cooling
Image Electrochemical decontamination of electrically conducting surfaces „EDeKo II“
Image Low Temperature Measuring Service
Image Influenced melting point of water by magnetic field
Image Computational fluid dynamics CFD
Image Tensile and compression testing
Image Multifunctional electronic modules for cryogenic applications
Image Mass Spectrometer
Image Breakthrough Sensor for Adsorption Filters (BelA)

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary

Your Request