Current research projects

Image Multifunctional electronic modules for cryogenic applications
Image Software for technical building equipment
Image Panel with indirect evaporative cooling via membrane
Image Lifetime prediction of hermetic compressor systems
Image Innovative Parahydrogen Generator Based on Magnets
Image Brine (water)-water heat pump
Image Behavior of multiphase cryogenic fluids
Image Investigation of coolants
Image Development of test methods and test rigs for stationary integrated refrigeration units
Image Certifiable connection types in cryogenics
Image Cold meter
Image Service offer for Leak Detection and Tightness Test
Image Certification of efficient air conditioning and ventilation systems through the new "indoor air quality seal" for non-residential buildings
Image ZeroHeatPump
Image Refrigerants, lubricants and mixtures
Image Low temperature – test facilities

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary