Current research projects

Image Intelligent innovative power supply for superconducting coils
Image Lifetime prediction of hermetic compressor systems
Image Low Temperature Measuring Service
Image Non- invasive flow measurements
Image Test procedures for electrical components
Image Pulse-Tube-Refrigerator with sealed compressor
Image Modular storage system for solar cooling
Image Refrigerants, lubricants and mixtures
Image High temperature heat pump
Image Influenced melting point of water by magnetic field
Image Ice Slurry Generation
Image ZeroHeatPump
Image In-Situ-Swelling Behaviour of Polymer Materials in Flammable Fluids
Image Test rigs for refrigeration and heat pump technology
Image Characterisation of Superconductors in Hydrogen Atmosphere
Image Humidifier System for High-Purity Gases

You are here:  Home /  Measurements and Tests


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary