Current research projects

Image Testing of mobile leak detectors according to DIN EN 14624
Image Reduction of primary noise sources of fans
Image Optimizing HVAC operation with machine learning
Image Development of test methods and test rigs for stationary integrated refrigeration units
Image Lifetime prediction of hermetic compressor systems
Image Non- invasive flow measurements
Image CFE-Test of Cooker Hoods
Image Air-flow test rig for fan characteristic measurement
Image Corrosion inhibitor for ammonia absorption systems
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image Performance tests of refrigerant compressors
Image Panel with indirect evaporative cooling via membrane
Image Cryogenic liquid piston pumps for cold liquefied gases like LIN, LOX, LHe, LH2, LNG, LAr
Image Cryostats, Non-Metallic and Metallic
Image Software modules
Image Solar Cooling

You are here:  Home /  Measurements and Tests


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary