Current research projects

Image Preformance measurements of heat exchangers
Image Test method for high - temperature heat pump - oils
Image Development of a Cryogenic Magnetic Air Separation Unit
Image Hybrid- Fluid for CO2-Sublimation Cycle
Image Calibration leak for the water bath leak test
Image In-Situ-Swelling Behaviour of Polymer Materials in Flammable Fluids
Image Filter Tests
Image Low temperature – test facilities
Image Performance tests of refrigerant compressors
Image Intelligent innovative power supply for superconducting coils
Image Swirl-free on the move...
Image Breakthrough Sensor for Adsorption Filters (BelA)
Image Thermal engines
Image Low noise and non metallic liquid-helium cryostat
Image Certification of efficient air conditioning and ventilation systems through the new "indoor air quality seal" for non-residential buildings
Image Refrigerants, lubricants and mixtures

You are here:  Home /  Measurements and Tests


Mass Spectrometer

Industrie und Forschungsinstitute

Gunar Schroeder

+49-351-4081-5129

Determining the composition of gas mixtures in the high or ultra-high vacuum range

Mass spectrometry is a method for determining the composition of gas mixtures in the high or ultra-high vacuum range. It can be used to carry out qualitative and quantitative gas analyses and leak detection or to determine trace contamination.

The individual neutral gas particles are ionized by an electron impact ion source and separated from each other in an electric quadrupole field according to their mass-to-charge ratio. The filtered ions are then recorded by a detector. The ion current detected is proportional to the partial pressure of the respective gas component. For a more precise quantitative gas analysis, the relatively operating measuring instrument can be calibrated with a calibration gas.

The test set-up consists of a test chamber with a connected turbo-molecular-pump and fore vacuum pump. The mass spectrometer requires a vacuum <1,5×10-7 psi (<1×10-5 mbar) to be able to measure with high precision. The mass spectrometer and a vacuum sensor are mounted on the test chamber. The mass spectrometer is connected to a measuring PC with corresponding software.

Available measurement modes are:

  • Vacuum analysis, to determine residual gas components, evaluation as a plot of ion current and mass
  • Leak detection, with helium or another tracer gas, the ion current of the tracer gas or the corresponding mass is recorded and plotted over the time
  • Mass spectrum, evaluation of ion current and mass with user-defined specifications and time trend
Paramter Test value
Analysis Vacuum analysis, leak detection, mass spectrum, trace contamination
Detectable gases H2, He, N2, H2O, O2, CO2 usw.
Mass range 1 – 100 Da (1,66×10-27 – 1,66×10-25 kg)
Detection limit 1.5×10-15 psi / 3×10-13 mbar
Min. vacuum pressure and max. temperature range <7.3×10-6 psi / 5×10-4 mbar, < 150°C / 302°F

[Translate to EN:]


Your Request

Further Projects - Measurements and Tests

Image

Refrigerants, lubricants and mixtures

Determination of working fluid properties

Image

Measurements on ceiling mounted cooling systems

Comparative performance measurement

Image

Solar Cooling

Solar Cooling with Photovoltaic

Image

Hydrogen and methane testing field at the ILK

Simultaneously pressures up to 1,000 bar, temperatures down to –253°C

Image

Calibration of Low Temperature Sensors

According to the comparative measurement method