Current research projects

Image Energy efficiency consulting - cogeneration systems
Image Reduction of primary noise sources of fans
Image Hydrogen and methane testing field at the ILK
Image Innovative small helium liquefier
Image Innovative Parahydrogen Generator Based on Magnets
Image Micro heat exchangers in refrigeration
Image Electrochemical decontamination of electrically conducting surfaces „EDeKo II“
Image Investigation of coolants
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image Low Temperature Measuring Service
Image Software for test rigs
Image Micro fluidic expansion valve
Image Industry 4.0 membrane heat and mass exchanger (i-MWÜ4.0)
Image Range of services laboratory analyses
Image Test method for high - temperature heat pump - oils
Image CO₂ GAS HYDRATES FOR SUSTAINABLE ENERGY AND COOLING SOLUTIONS

You are here:  Home /  Research and Development


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary

Your Request

Further Projects - Research and Development

Image

High Capacity Pulse Tube Cooler

for Cryogenic High-Power Applications

Image

Thermal engines

Power Generation from Waste Heat

Image

Helium extraction from natural gas

Innovative solutions for helium extraction

Image

Ice Slurry Generation

Using Direct Evaporation

Image

Pulse-Tube-Refrigerator with sealed compressor

for mobil use in the hydrogen technology