Current research projects

Image Verification of storage suitability of cryo tubes
Image Test method for high - temperature heat pump - oils
Image Innovative Manufacturing Technologies for Cryosorption Systems
Image Tribological investigations of oil-refrigerant-material-systems
Image Computational fluid dynamics CFD
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image Testzentrum PLWP at ILK Dresden
Image Non- invasive flow measurements
Image Air-flow test rig for fan characteristic measurement
Image Hybrid- Fluid for CO2-Sublimation Cycle
Image Pulse-Tube-Refrigerator with sealed compressor
Image Electrochemical decontamination of electrically conducting surfaces „EDeKo II“
Image Innovative small helium liquefier
Image Influenced melting point of water by magnetic field
Image Preformance measurements of heat exchangers
Image Multifunctional electronic modules for cryogenic applications

You are here:   /  Home


Mass Spectrometer

Industrie und Forschungsinstitute

Gunar Schroeder

+49-351-4081-5129

Determining the composition of gas mixtures in the high or ultra-high vacuum range

Mass spectrometry is a method for determining the composition of gas mixtures in the high or ultra-high vacuum range. It can be used to carry out qualitative and quantitative gas analyses and leak detection or to determine trace contamination.

The individual neutral gas particles are ionized by an electron impact ion source and separated from each other in an electric quadrupole field according to their mass-to-charge ratio. The filtered ions are then recorded by a detector. The ion current detected is proportional to the partial pressure of the respective gas component. For a more precise quantitative gas analysis, the relatively operating measuring instrument can be calibrated with a calibration gas.

The test set-up consists of a test chamber with a connected turbo-molecular-pump and fore vacuum pump. The mass spectrometer requires a vacuum <1,5×10-7 psi (<1×10-5 mbar) to be able to measure with high precision. The mass spectrometer and a vacuum sensor are mounted on the test chamber. The mass spectrometer is connected to a measuring PC with corresponding software.

Available measurement modes are:

  • Vacuum analysis, to determine residual gas components, evaluation as a plot of ion current and mass
  • Leak detection, with helium or another tracer gas, the ion current of the tracer gas or the corresponding mass is recorded and plotted over the time
  • Mass spectrum, evaluation of ion current and mass with user-defined specifications and time trend
Paramter Test value
Analysis Vacuum analysis, leak detection, mass spectrum, trace contamination
Detectable gases H2, He, N2, H2O, O2, CO2 usw.
Mass range 1 – 100 Da (1,66×10-27 – 1,66×10-25 kg)
Detection limit 1.5×10-15 psi / 3×10-13 mbar
Min. vacuum pressure and max. temperature range <7.3×10-6 psi / 5×10-4 mbar, < 150°C / 302°F