Current research projects

Image Cool Up
Image Characterisation of Superconductors in Hydrogen Atmosphere
Image Calibration of Low Temperature Sensors
Image All-in-one device for freeze-drying and production of biomaterial
Image Testing of mobile leak detectors according to DIN EN 14624
Image Service offer for Leak Detection and Tightness Test
Image Cryogenic liquid piston pumps for cold liquefied gases like LIN, LOX, LHe, LH2, LNG, LAr
Image Humidifier System for High-Purity Gases
Image Filter Tests
Image Calibration leak for the water bath leak test
Image Micro heat exchangers in refrigeration
Image Certifiable connection types in cryogenics
Image Refrigerants, lubricants and mixtures
Image Brine (water)-water heat pump
Image Electrical components in refrigeration circuits
Image Preformance measurements of heat exchangers

You are here:  Home /  Measurements and Tests


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary

Your Request

Further Projects - Measurements and Tests

Image

Test procedures for electrical components

Insulating properties of hermetic compressors

Image

3D - Air flow sensor

Anemometer for determination of 3d-air flow