Current research projects

Image Performance tests of condensing units
Image Heat2Power
Image Overall System Optimization of Refrigeration Plant Systems for Energy Transition and Climate Protection
Image Behavior of multiphase cryogenic fluids
Image Performance tests of refrigerant compressors
Image Multifunctional electronic modules for cryogenic applications
Image Production of novel barrier layers on polymer materials to reduce hydrogen permeation
Image Innovative small helium liquefier
Image Test method for high - temperature heat pump - oils
Image Low temperature – test facilities
Image Innovative Parahydrogen Generator Based on Magnets
Image High temperature heat pump
Image Laseroptical measurement
Image Hybrid- Fluid for CO2-Sublimation Cycle
Image Calibration of Low Temperature Sensors
Image Software for technical building equipment

You are here:  Home /  Research and Development


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary