Current research projects

Image Software for test rigs
Image Investigation of materials
Image Certification of efficient air conditioning and ventilation systems through the new "indoor air quality seal" for non-residential buildings
Image Laseroptical measurement
Image Ice Slurry Generation
Image In-situ investigation concerning the swelling behaviour of polymer materials under elevated pressures and temperatures
Image Development of test methods and test rigs for stationary integrated refrigeration units
Image Testzentrum PLWP at ILK Dresden
Image Multifunctional electronic modules for cryogenic applications
Image Investigation of coolants
Image Thermostatic Expansion Valves
Image ZeroHeatPump
Image Influenced melting point of water by magnetic field
Image Innovative Manufacturing Technologies for Cryosorption Systems
Image Testing of mobile leak detectors according to DIN EN 14624
Image Helium extraction from natural gas

You are here:  Home /  Research and Development


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary