Aktuelle Forschungsprojekte

Image Industry 4.0 membrane heat and mass exchanger (i-MWÜ4.0)
Image Testing of mobile leak detectors according to DIN EN 14624
Image Corrosion inhibitor for ammonia absorption systems
Image Innovative small helium liquefier
Image Non- invasive flow measurements
Image IO-Scan - Integral measuring optical scanning method
Image Certification of efficient air conditioning and ventilation systems through the new "indoor air quality seal" for non-residential buildings
Image Investigation according to DIN EN ISO 14903
Image Hybrid- Fluid for CO2-Sublimation Cycle
Image Testzentrum PLWP at ILK Dresden
Image Thermostatic Expansion Valves
Image Range of services laboratory analyses
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image Behavior of multiphase cryogenic fluids
Image Optimizing HVAC operation with machine learning
Image Laseroptical measurement

You are here:  Home /  Research and Development


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary