Current research projects

Image Mass Spectrometer
Image Low noise and non metallic liquid-helium cryostat
Image Swirl-free on the move...
Image Reducing the filling quantity
Image Air-water heat pumps
Image Micro heat exchangers in refrigeration
Image Innovative cryogenic cooling system for the recondensation / liquefaction of technical gases up to 77 K
Image Service offer for Leak Detection and Tightness Test
Image Thermostatic Expansion Valves
Image Behavior of multiphase cryogenic fluids
Image Verification of storage suitability of cryo tubes
Image Investigation of materials
Image Certification of efficient air conditioning and ventilation systems through the new "indoor air quality seal" for non-residential buildings
Image In-Situ-Swelling Behaviour of Polymer Materials in Flammable Fluids
Image Performance tests of condensing units
Image Development of a Cryogenic Magnetic Air Separation Unit

You are here:  Home /  Testing Equipment


Humidifier System for High-Purity Gases

Industry and Research Instituts

Moritz Kuhn

+49-351-4081-5120

Nafion - moisture transfer

At the ILK Dresden a humidification system was developed and manufactured which is especially suitable for research facilities where very low dew point temperatures have to be adjusted and at the same time a high gas purity has to be guaranteed. The core component is a PermaPure® humidifier with Nafion® moisture exchange. The humidification system precisely controls the dew point via a bypass mixing with dry gas and also takes over the automatic filling of evacuated containers. The control system is optimised to prevent recondensation. Humidification can be controlled from any plant system via an interface. Table 1 shows the technical data of a humidification system for the humidification of synthetic air realized for the KIT in Karlsruhe. The system can be adapted for other applications, and all parameters can be adjusted to the respective requirements.

Table  1: Technical Data Humidifier System
Parameter Value
setpoint gas flow 750 litre/min
maximum relative humidity 95%
minimum dew point -60°C / -76°F
degree of gas purity 5.0
gas purity, number of particle < 5 particle/cm³