Current research projects

Image Performance tests of refrigerant compressors
Image Cryogenic liquid piston pumps for cold liquefied gases like LIN, LOX, LHe, LH2, LNG, LAr
Image Test procedures for electrical components
Image Characterisation of Superconductors in Hydrogen Atmosphere
Image Modular storage system for solar cooling
Image Reduction of primary noise sources of fans
Image Ionocaloric cooling
Image Innovative Parahydrogen Generator Based on Magnets
Image Overall System Optimization of Refrigeration Plant Systems for Energy Transition and Climate Protection
Image Development of a Cryogenic Magnetic Air Separation Unit
Image Lifetime prediction of hermetic compressor systems
Image Software for technical building equipment
Image Thermal engines
Image Investigation according to DIN EN ISO 14903
Image Hybrid- Fluid for CO2-Sublimation Cycle
Image Test method for high - temperature heat pump - oils

You are here:   /  Home


Calibration of Low Temperature Sensors

Industry and R&D

Dr. rer. nat. Matthias Schneider

+49-351-4081-5126

According to the comparative measurement method

Accurate and reliable measurements are the requirements for high-quality research and development or industrial production. Using the comparative measurement method the following sensors can be calibrated:

  • Resistance Temperature Detectors (RTD) with positive temperature coefficient (like Platinum or Rhodium-Iron RTDs)
  • RTD with negative temperature coefficient (like Germanium or Cernox™ RTDs)
  • Diode and thermocouples.

With an automatic measuring station several temperatures according to a sample point table will be adjusted. The calibration is carried out with secondary thermometers that are calibrated by an authority. Furthermore a fix point check with LHe or LN2 will be performed. In addition, a calbriated Cernox sensor is available for measurements in the low temperature range. The measuring facility allows the calibration of about 30 sensors at once. There are continuous extensions for different sensor types.

Parameter

IndicatorMinimalMaximal
Measuring range1.5 K350 K
Measuring accuracy5 mK (1.5 ... 4.2 K)50 mK (80 ... 350 K)
Sample dimensioncommon sensor packages
Sample preparationnot necessary